8100 : Entering Calibration Mode

Modified on Tue, 07 Sep 2021 at 03:56 PM

The 8100 Advanced Multimeter can be fully adjusted from the front panel, or over the remote interface.

To enter Calibration Mode on the 8100 Series via the front panel, press the 'MENU' key, then select 'Calibration Menu' from the Main Menu

Enter the password (default 1234) followed by pressing the ENTER softkey.


Enter the calibration Menu and then select 'CALIBRATION MODE [OFF]'

If a temperature warning message appears, leave the multimeter to warm up for longer, as this is indicating that the multimeter is not yet thermally stable at the normal operation temperature.

When the menu item displays 'CALIBRATION MODE [ON]', press the BACK or EXIT key to return to main operation.

The BLUE shift key indicates that calibration mode is presently active, along with the RED C in the bottom right of the screen


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select atleast one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article